(EDS Spectrum showing elemental peak height as a function of X-ray counts collected (LC) Elements Identified Major (> 10%): oxygen, carbon, calcium, phosphorus Minor (1-10%): NA Trace (< 1%): magnesium Area X-ray counts collected: 1,051,106 Live Tim...) |
|||
Line 2: | Line 2: | ||
Elements Identified | Elements Identified | ||
− | Major (> 10%): oxygen, carbon, calcium, phosphorus | + | Major (> 10%): oxygen, carbon, calcium, phosphorus. |
− | Minor (1-10%): NA | + | Minor (1-10%): NA. |
− | Trace (< 1%): magnesium | + | Trace (< 1%): magnesium. |
− | Area X-ray counts collected: 1,051,106 | + | Area X-ray counts collected: 1,051,106. |
− | Live Time: 30.8 seconds | + | Live Time: 30.8 seconds. |
− | Magnification: 500x | + | Magnification: 500x. |
Analyzed using Oxford X-Max80 energy-dispersive X-Ray spectroscopy (EDS) and Oxford AZtec platform software. Pigment sample was applied to carbon tape and mounted onto aluminum examination stub. X-rays from sample were excited by a 20 kV SEM electron beam in a low vacuum chamber (10 Pascal). The resolution of the X-Max80 silicon drift detector is 123 eV at MnKα wavelength. | Analyzed using Oxford X-Max80 energy-dispersive X-Ray spectroscopy (EDS) and Oxford AZtec platform software. Pigment sample was applied to carbon tape and mounted onto aluminum examination stub. X-rays from sample were excited by a 20 kV SEM electron beam in a low vacuum chamber (10 Pascal). The resolution of the X-Max80 silicon drift detector is 123 eV at MnKα wavelength. |
Latest revision as of 09:11, 10 January 2017
EDS Spectrum showing elemental peak height as a function of X-ray counts collected (LC)
Elements Identified Major (> 10%): oxygen, carbon, calcium, phosphorus. Minor (1-10%): NA. Trace (< 1%): magnesium.
Area X-ray counts collected: 1,051,106. Live Time: 30.8 seconds. Magnification: 500x. Analyzed using Oxford X-Max80 energy-dispersive X-Ray spectroscopy (EDS) and Oxford AZtec platform software. Pigment sample was applied to carbon tape and mounted onto aluminum examination stub. X-rays from sample were excited by a 20 kV SEM electron beam in a low vacuum chamber (10 Pascal). The resolution of the X-Max80 silicon drift detector is 123 eV at MnKα wavelength.
File history
Click on a date/time to view the file as it appeared at that time.
Date/Time | Thumbnail | Dimensions | User | Comment | |
---|---|---|---|---|---|
current | 12:53, 4 January 2017 | 1,460 × 441 (65 KB) | MKullman (talk | contribs) | EDS Spectrum showing elemental peak height as a function of X-ray counts collected (LC) Elements Identified Major (> 10%): oxygen, carbon, calcium, phosphorus Minor (1-10%): NA Trace (< 1%): magnesium Area X-ray counts collected: 1,051,106 Live Tim... |