Yellowtops (Flaveria haumanii) LC
Description
Historical importance
Summary of results
Analytical instrumentation and procedures
HPLC-DAD-MS analysis was performed with an Agilent 1100 liquid chromatography system consisting of an automatic injector, a gradient pump, a HP series 1100 DAD, and an Agilent series 1100 VL on-line atmospheric pressure ionization electrospray ionization mass spectrometer. Separations were done on a Vydac 214TP52 analytical column (2.1 mm diameterX250 mm; 5-ím particle size). The column was eluted at a flow rate of 0.2 mL/min with a tertiary gradient of water (A),acetonitrile (B), and 1% (v/v) aqueous formic acid (C) with the following elution program: 0 min, 90% A, 5% B, 5% C; 0-55 min, a linear gradient to 35% A, 60% B, 5% C; 55-60 min, a linear gradient elution to 15% A, 80% B, 5% C; 60-62 min, isocratic elution at 15% A, 80% B, 5% C; 62-70 min gradient elution to 90% A, 5% B, 5% C; and reequilibration with the latter solvent for 15 min. The mass spectrometer was run both in the negative and positive ion mode.
Chromatograms
Sample Information
Identified compounds
Compound | RT (min.) | MW | UV/vis | Other |
---|---|---|---|---|
quercetin 3-O-sulfate | 24.4 | 382 | 278,348 | Comments here |
isorhamnetin 3-O-sulfate | 27.1 | 394 | 280,350 | |
kaempferol | 39.0 | 286 | 280,395,368 | |
isorhamnetin | 39.0 | 316 | 280,395,368 |